Calibration<br />

Step Operator Action Scrolling Marquee Main Display Notes<br />

23 Press to accept the new slope and<br />

E 0 values<br />

or<br />

Press to abort the calibration. The<br />

new slope and E 0 values will not be saved.<br />

The last saved calibration prior to the<br />

aborted calibration will be used in the<br />

measure mode.<br />

24 Pull the diverter valve out.<br />

Turn the inlet valve on.<br />

Close the calibration port on<br />

the flow cell cap.<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

NEW CALIBRATION SLOPE AND E 0<br />

PRESS ENTER TO ACCEPT OR PRESS<br />

EXIT TO ABORT WITHOUT SAVING<br />

PULL DIVERTER VALVE OUT<br />

TURN INLET VALVE ON<br />

CLOSE CAL PORT<br />

PRESS ENTER<br />

25 Press . PULL DIVERTER VALVE OUT<br />

TURN INLET VALVE ON<br />

CLOSE CAL PORT<br />

PRESS ENTER<br />

Calculated Slope<br />

and E 0<br />

FLO On<br />

FLO On<br />

26 System will proceed to measure mode. Actual temperature reading Actual concentration<br />

reading<br />

If is pressed,<br />

refer to the<br />

Calibration Abort<br />

Steps section.<br />

System enters<br />

measure mode, but<br />

will remain on hold<br />

until the hold timer<br />

expires. Default hold<br />

time is 30 minutes.<br />

IV-9