<strong>Analyzer</strong> Maintenance<br />

valve stem<br />

O-rings (2)<br />

Yearly Preventive<br />

Maintenance<br />

Electrodes<br />

Diverter Valve O-rings<br />

valve stem<br />

retaining ring<br />

manifold<br />

face<br />

Figure V-1<br />

Flow Cell Assembly<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

1. Replace the Thermo Scientific <strong>Orion</strong> reference electrode<br />

(Cat. No. 100057). Refer to Chapter II, Installing a New Reference<br />

Electrode for step-by-step instructions.<br />

2. Replace the Thermo Scientific <strong>Orion</strong> chloride sensing electrode<br />

(Cat. No. 210017). Refer to Chapter II, Installing a New <strong>Chloride</strong><br />

Electrode for step-by-step instructions.<br />

3. Replace the calibration solutions (Cat. No. 211740HL)<br />

Dispose of all old solutions according to the MSDS instruction. MSDS<br />

files can be downloaded at www.thermo.com/processwater.<br />

Valve Stem O-rings<br />

1. Remove the two screws that hold the valve stem assembly in place.<br />

2. Gently pull on the valve stem to remove.<br />

3. Slide off the two old O-rings and replace them. It is not necessary to<br />

remove the retaining ring.<br />

4. Place the assembly back into the manifold block and tighten screws.<br />

Note: Do not use O-ring grease.<br />

Manifold Face O-rings<br />

1. Remove the flow cell manifold assembly.<br />

2. Remove the three O-rings from the face of the flow cell manifold.<br />

3. Insert the new O-rings on the face of the flow cell manifold.<br />

4. Replace the flow cell manifold assembly.<br />

V-4