<strong>Analyzer</strong> Operation<br />

Shutdown and<br />

Start-Up Procedure<br />

Shutdown<br />

Air Pump Shutdown<br />

<strong>Analyzer</strong> Shutdown<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

The following steps should be taken if a loss of sample flow is expected for<br />

more than one day.<br />

Warning: Be sure to read and observe the following requirements.<br />

Danger: Turn off the air pump prior to removing the reference electrode<br />

filling solution bottle. The air pump will spatter filling solution outward as<br />

the bottle is removed.<br />

1. To access air pump, press .<br />

2. Press / to scroll though parameters until TEST appears in<br />

marquee window.<br />

3. Press to access the TEST menu.<br />

4. Press / to scroll through the TEST menu until AIr appears<br />

in the main display.<br />

5. Press .<br />

6. Press / to toggle between On and OFF settings for the air<br />

pump power. Select the OFF setting. Changing the air pump setting to<br />

On or OFF will be indicated by a click sound.<br />

7. Press to accept the setting and return to the TEST menu.<br />

8. Press to return to the measure mode.<br />

1. Shut off sample flow prior to the analyzer inlet. Refer to Chapter IV,<br />

Flow Regulation.<br />

2. Disconnect the power. Refer to Chapter II, Wiring the <strong>Analyzer</strong>.<br />

3. Drain the flow cell.<br />

4. Carefully pull the chloride and reference electrodes out of the top of the<br />

flow cell and let them hang by their connectors.<br />

5. Locate the protective cap from reference electrode kit and place it on<br />

the base of the reference electrode to prevent it from drying out.<br />

III-35