<strong>Analyzer</strong> Preparation<br />

Installation of a New<br />

<strong>Chloride</strong> Electrode<br />

Figure II-8<br />

Flow Cell Cap<br />

Installation of the<br />

ATC Probe<br />

Thermo Scientific <strong>Orion</strong> <strong>2117HL</strong> <strong>High</strong> <strong>Level</strong> <strong>Chloride</strong> <strong>Analyzer</strong> <strong>User</strong> <strong>Guide</strong><br />

The Thermo Scientific <strong>Orion</strong> chloride electrode (Cat. No. 210017)<br />

must be used in conjunction with the Thermo Scientific <strong>Orion</strong> reference<br />

electrode (Cat. No. 100057).<br />

1. Unpack the chloride electrode (Cat. No. 210017) and carefully remove<br />

the protective cap. Save the cap for future storage of the electrode.<br />

Use the polishing strip (Cat. No. 948201) to gently polish the sensing<br />

surface of the chloride electrode for about 30 seconds.<br />

2. Insert the chloride electrode into its port in the flow cell cap. Refer to<br />

Figure II-8 for the location.<br />

3. Plug the electrode cable marked “Sensing” into the top of the electrode.<br />

Be sure to push back the black cap to verify a secure connection<br />

between the male and female pin connection prior to tightening.<br />

4. Tighten the screw cap connection to the cable.<br />

Note: Do not twist the cable while tightening the connection. Twisting<br />

may cause damage requiring premature replacement of the cable.<br />

5. Wait at least one hour before calibrating the analyzer.<br />

Sensor<br />

Reference<br />

Calibration Port<br />

The automatic temperature compensation (ATC) probe (Cat. No. 2100TP)<br />

is already connected to the correct terminal for temperature upon delivery.<br />

1. Insert the ATC probe into its port in the flow cell cap. Refer to Figure<br />

II-8 for the location.<br />

ATC<br />

II-11