UWE Bristol Engineering showcase 2015
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Stress (Pa)<br />
Stress (Pa)<br />
Tom Leggett<br />
BEng Mechanical <strong>Engineering</strong><br />
Design of a Prosthetic Ankle using Composite Materials<br />
Design Stage 1 – Basic shape and dimensions<br />
This design allowed flexibility in achieving the<br />
defined requirements of the prosthetic ankle. The<br />
main benefit is that there is no complicated pivot<br />
mechanism required to achieve dorsiflexion and<br />
plantar flexion. The curved shape acts as both a<br />
spring and a pivot for energy return and imitation<br />
of ankle mechanics.<br />
Design Stage 2 – Angles and Radii<br />
The basic design has been created, with some<br />
dimensions being constant and others being the<br />
variable. The table below indicated which<br />
dimensions are which:<br />
Constant Dimension<br />
Top Section Length (70mm)<br />
Variable Dimension<br />
Radius of Curve (r)<br />
Bottom Section Length (80mm) Angle of Top Section (θ 1 )<br />
Thickness (3.75mm) Angle of Bottom Section (θ 2 )<br />
Width (80mm)<br />
7.00E+06<br />
6.00E+06<br />
5.00E+06<br />
4.00E+06<br />
3.00E+06<br />
2.00E+06<br />
1.00E+06<br />
0.00E+00<br />
Outside Centreline Stress - Varying r<br />
a45b10r20 a45b10r30 a45b10r35<br />
0 50 100 150 200 250<br />
True Distance Along Path<br />
Design Stage 2 – Results<br />
35° produced less<br />
variation between the<br />
two peaks as well as a<br />
lower maximum value<br />
of 4.6 MPa. An angle of<br />
0° for the lower section<br />
would make it parallel<br />
to the floor<br />
encouraging flat foot to<br />
occur. It was therefore<br />
decided that an angle<br />
of 5° for θ 2 would be<br />
chosen for design stage<br />
2. It was found that the<br />
bending machine could<br />
only bend to a<br />
minimum diameter of<br />
50mm Therefore 25mm<br />
radius was chosen<br />
Design Stage 3 – Composite Layups<br />
Epoxy resin and E-glass were the chosen matrix<br />
and reinforcement for layup testing. 7 layups were<br />
tested with the aim of achieving stresses close as<br />
possible to that of the project objective.<br />
[0 5 /90 2 /90/90 2 /0 5 ] Layup 7 had a slightly lower<br />
stress over the load area due to the increase of 0°<br />
plies, but a much higher stress on the outside and<br />
inside centreline. The peak values were 3.05 MPa<br />
and 3.1 MPa for the outside and inside<br />
respectively. The stresses along the edges<br />
exhibited the same results but with a higher peak<br />
stress of 3.45 MPa for both the inside and outside.<br />
The outside centreline and edge strains mimicked<br />
the previous graph shapes but with a doubled<br />
peak value of 0.95E-03. The increased flexibility of<br />
layup 7 is summarised by the massive increase in<br />
deflection to 4.2mm under loading. The final<br />
results of layup 7 provide a successful design.<br />
4.00E+06<br />
3.50E+06<br />
3.00E+06<br />
2.50E+06<br />
2.00E+06<br />
1.50E+06<br />
1.00E+06<br />
5.00E+05<br />
0.00E+00<br />
Stress for Varying Layups - Outside Edge<br />
[0/0/0/45/-45/45/-45][0][-45/45/-45/45/0/0/0]<br />
[0/90/0/45/-45/0/0][0][0/0/-45/45/0/90/0]<br />
[0/0/0/0/0/90/90][90][90/90/0/0/0/0/0]<br />
0 50 100 150 200 250<br />
True Distance Along Path (mm)<br />
Final Product Manufacture and Testing<br />
The inside edge clearly shows signs of the fibres<br />
buckling, especially on the outside face of the<br />
laminate. The simulations and these results agree<br />
that the maximum compression occurs at the<br />
bottom of the curved section, at half way along<br />
the entire length of the model.<br />
Project Supervisor<br />
Dr Ruth Jones<br />
Project summary<br />
An investigation into the design of a prosthetic ankle<br />
using composite materials to mimic the<br />
characteristics of a normal human was undertaken.<br />
The minimum load and maximum stress experienced<br />
by a normal ankle was found from literature. Current<br />
designs of prosthetic feet were used to create an<br />
initial model on Abaqus, a finite element analysis<br />
software. The best design was chosen and then<br />
manufactured in the University West of England<br />
laboratory. The simulation results were then verified<br />
by testing the completed composite ankle model in a<br />
compression test. The model withstood the minimum<br />
load and hence verified the design of the ankle. The<br />
design is not the complete solution for a prosthetic<br />
foot, but provided a basis by which the other<br />
components of the foot can be designed around.<br />
Project Objectives<br />
Design a prosthetic ankle capable of supporting the<br />
weight of an average human during a normal walking<br />
gait cycle, whilst replicating the stress of a normal<br />
ankle joint.<br />
• Minimum load 3500N<br />
• Maximum stress 3.464 MPa<br />
Project Conclusion<br />
The modelling on Abaqus found the best angles for θ 1<br />
and θ 2 were 35° and 5° respectively. Both were found<br />
to have the lowest peak stress and less variation<br />
between the two peaks on the top and bottom of the<br />
curved section. A smaller radius provided less stress<br />
and strain, but due to manufacturing limitations a<br />
radius of 25mm was selected. A layup of<br />
[0 5 /90 2 /90/90 2 /0 5 ] was the final orientation chosen,<br />
as it gave a maximum stress value of 3.45 MPa,<br />
extremely close to that defined in the specification.<br />
The final model testing was a simplistic compression<br />
test, but it proved successful. The ankle passed the<br />
minimum load requirement of 3500N and began to<br />
show signs of failure at 4700N. The difference<br />
between the simulation and the real world test<br />
highlighted issues with the model simplification, but<br />
despite this a successful product was made that met<br />
the specification and aims of the investigation.