The +10V signal should read as follows:<br />

Minimum Nominal Maximum<br />

+9.990V +10.000V +10.010V<br />

Performance Verification<br />

4. Check the +15 volt supply by placing one probe of your digital multimeter<br />

on pin 24 (+15V) of the 78-pin connector. Place the other probe on pin 1<br />

(ground). Both pin locations are shown in Figure A-4.<br />

WARNING: Make sure the probes are on the correct connector pins.<br />

The +15V signal should read as follows:<br />

Minimum Nominal Maximum<br />

+14.25V +15.00V +15.75V<br />

You are now finished checking the calibration of your programmer. If the<br />

reference voltages and master clock match the listed specifications, your<br />

programmer is operating within factory specifications.<br />

Reassembling the Programmer<br />

Follow the steps below to reassemble your programmer:<br />

1. Set the top cover back onto the programmer.<br />

2. Using a #1 Phillips screwdriver, replace the two (or three) screws you<br />

removed from the rear panel in step 4. See Figure A-1 for the location of<br />

the screw holes.<br />

3. Using a #1 Phillips screwdriver, replace the three screws you removed from<br />

the top of the programmer in step 6. See Figure A-2 for the location of the<br />

screw holes.<br />

Your programmer is reassembled and ready for use.<br />

<strong>3980xpi</strong>/3980/3900/2900 User <strong>Manual</strong> A-5