Spot Color Simulation<br />

Step 5: Add Spot Colors<br />

http://www.<strong>image2output</strong>.com<br />

// WEB«<br />

1. In the SPOT COLOR CORRECTION dialog (Fig. 15-6), click on NEW SPOT<br />

COLORS. The MANUAL SPOT COLOR INPUT dialog appears (Fig. 15-7).<br />

2. Enter a name for the new color into the COLOR NAME box.<br />

3. Enable the AUTO ADD option if the measured colors should be automatically<br />

added to the list in the SPOT COLOR CORRECTION dialog.<br />

Otherwise you have to click the ADD button after each measurement.<br />

4. Specify the TONAL VALUES to be measured.<br />

Enter 100 to measure solid tones only. (Instead of measuring, you<br />

could also type the Lab values into the L a b boxes on the left.)<br />

Enter 100 60 40, for example, to additionally measure the 60% and<br />

40% value. The values can be separated by any character which is<br />

no digit. You can also select from a list of predefined tonal values.<br />

5. If known, specify the spot color’s OPACITY and PRINT ORDER.<br />

6. The selected measurement device is indicated in the dialog. Click the<br />

SELECT button if you want to select another device.<br />

7. Click the START MEASUREMENTS button.<br />

8. Usually the following steps need to be taken for the first measurement<br />

only:<br />

• A message appears telling you to place the instrument onto its<br />

base. A white calibration will be performed automatically.<br />

• A message appears telling you to place the instrument onto the<br />

paper white (a white area of the proofing substrate or sample).<br />

9. Now the button you have clicked indicates MEASURE 100%.<br />

Place the measurement head on the patch with the solid tone. Then<br />

press the button on the measurement device.<br />

10.When measuring spot color halftones, the button indicates the next<br />

tonal value to be measured. Measure this patch in the same way.<br />

11.Having measured all tonal values, the color name and the measurement<br />

values appear automatically in the MANUAL SPOT COLOR INPUT<br />

dialog (see Fig. 15-7). If the AUTO ADD option is not enabled, however,<br />

you have to click the ADD button.<br />