17.06.2013 Views

FIFTH CANADIAN CONFERENCE ON NONDESTRUCTIVE ... - IAEA

FIFTH CANADIAN CONFERENCE ON NONDESTRUCTIVE ... - IAEA

FIFTH CANADIAN CONFERENCE ON NONDESTRUCTIVE ... - IAEA

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

- 224 -<br />

on the fringes. (The sensitivity setting on the instrument was very low, hence<br />

the defect would show even larger with the sensitivity increased to normal<br />

testing levels.)<br />

4. VERIFICATI<strong>ON</strong><br />

ïhe position of the cracks depends on the design of the turbine. Depending on<br />

the model, the experienced operator could tell ahead where the cracks will be<br />

found. That means that a probe can be designed with the highest sensitivity<br />

at the location of the anticipated cracks. Nevertheless, one needs a probe<br />

which covers a reasonably wide swath, in case defects with another orientation<br />

are present which ought to be detected.<br />

Test blocks were manufactured from homogeneous teeth by Electro Discharge<br />

Machining small artificial defects in the grooves (Figure 9)• One groove contained<br />

notches of different depths for calibration purposes, another groove<br />

contained identical notches which were offset to determine field-of-view of<br />

the ET probes.<br />

It was found before, that absolute probes have advantages when only one phenomenon<br />

is sought. 1 An absolute probe was made with both a wide and a limited-length<br />

field of view. A once-through test with this probe on the reference<br />

notches with respect to their position is shown in Figure 10.<br />

By changing the frequency, the optimum response was found, and it was not surprising<br />

that it was close to the design frequency. It was expected that a<br />

change in ratio of signals from small to large reference notches would also be<br />

found with a change of frequency but it was not (Figure 11).<br />

Since notch 2 is supposed to be exactly 0.2 mm deep and notch 3, 0.4 mm deep,<br />

the ideal signal ratio would be two. From 800 to 900 kHz the ratio was 2.3,<br />

considered the best obtainable. It was also decided that it was probably better<br />

to use a firm signal as a set-up rather than a smaller but noisier one and<br />

the 0.4 mm deep notch was a tentative choice.<br />

At 800 kHz a series of measurements with different instrument sensitivities<br />

was made (Figure 12). This showed that without increasing the noise level,<br />

signals could be boosted to show the 0.1 mm notch (No. 4, Figure 9) clearly.<br />

With this level of sensitivity, any dangerous crack would cause the indication<br />

at the lower sensitivity. The latter appears better for testing since too<br />

many indications would go off-screen if the sensitivity were too high. With a<br />

reduced sensitivity test may be more realistic and could be used to as close<br />

as 5 mm from the end. (Notches are at 20, 40, 60 and 78 mm from the left<br />

edge, Figure 13.) With this setting, the real crack in Figure 13 showed a<br />

start in steps but is fully established at 50 mm and carries on to the end.<br />

(The microscope showed it to fade at 4 mm from the end or 87 mm from the left<br />

which is revealed in the picture although not clearly.)<br />

1. G. Van Drunen and V.S. Cecco, CSNDT Journal, Vol. 5 No. 1, 1983 October

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!