- 362 -<br />

This is a reasonable approximation if t(n,m) is the average value<br />

of the depth over a circular area. The sum of the responses over<br />

the six nearest neighbours of the location that is to be<br />

investigated (n,,m) is calculated overt<br />

F(n,m) = s(n-1,m) + s(n, m-1) + s(n, m+1) + s(n-1, m-1)<br />

This can be expressed as<br />

+ s(n+1, m-1) + s(n+1,m)<br />

F(n,m) = 6t(n,m) + 2 t(n-1,m) + 2t(n,m-l)<br />

+ 2t(n+l,m) + 2t(n,m+l) + 2t(n-1,m-1)<br />

+ 2t(n+T,m-1) + 2t(n+2,m) + 2t(n-2,m)<br />

+ 2t(n+1,m-2) + 2t{n+1,m+i) + 2t(n-1,m-2)<br />

+ 2t(n-1,m+1) + t(n-2,m-1) + t(n-2,m+1)<br />

+ t(n+2,m-1) + t(n+2,m+1) + t(n,m-2) + t(n,m+2)<br />

The first sufficient approximation is t(n,m) = F(n,m)<br />

6<br />

This solution can be used for an iterative procedure for the<br />

solution of n X m linear equations with n X m unknowns, thus<br />

completing the transformation of the S matrix into a T matrix.<br />

The values of the T matrix can then be color coded to produce an<br />

image with depth indications. The above methodology applies both<br />

to wire wound and printed circuit coils. This procedure, which<br />

plots depth estimates at the center of each individual coil<br />

core, can also be used in the dynamic case.<br />

TESTS PERFORMED WITH THE TSI ARRAY<br />

This array was prepared in house. In short, twelve coils were<br />

arranged in a rectangular pattern of three by four coils, with<br />

the distance between the nearest neighbours being 1.75 nm - The<br />

individual coils serve both as driver and receiver.<br />

The performance of individual coils was assessed, namely the<br />

sensitivity to detect the shallow (.1mm deep by 2mm long)<br />

slot (Figure 8). The coils picked up the defect with no<br />

difficulties.<br />

This array was also mechanically scanned over an aluminum plate<br />

containing 3 3" hole. Figure 9 shows two isometric plots<br />

obtained by scanning the array over 33 m hole.<br />

We experienced some asymmetries: coils 1 to 6 had a larger left<br />

peak than right peak with the opposite being the case for ^ to<br />

12. This was due to the higher concentration of flux lines to