Breast supportPMMA slabBreast supportPMMA slabprevious locationPMMA slabA – mean ROI 1 DB – mean ROIC – SD ROI22(a)(b)FIG. 32. Set-up for (a) ghost creation image and (b) ghost measurement image.(10) Calculate the ghost image SDNR using the equation:Ghost image SDNR = A - BC(2)where the variables are as defined in Fig. 32(b).(11) View the ‘for processing’ version of the ghost measurement image on the radiologist review workstation.(12) Using a narrow window width (without exaggerating image noise) and an appropriate window level, inspectthe central part of the image where the boundary between the two areas of exposure lies. Record the windowwidth and window level on Chart 7 in Annex II.(13) Record the presence or absence of any visually observable ghost image on Chart 7 in Annex II.8.5.4.4. Interpretation of results and conclusionsTolerance:(1) Acceptable: Ghost image SDNR ≤2.0.(2) Alternative measure: When viewing the ghost measurement image with a typical, clinically used windowsetting, there should be no visible indication of the attenuator in the ghost creation image.8.5.4.5. Recommendations and corrective action(1) If the absolute value of the ghost image SDNR is more than the tolerance value, the service person should becontacted.(2) If ghosting is visible at clinically relevant window width and window level settings, the service person shouldbe contacted.8.5.4.6. Time frame for corrective actionImmediately: Corrective action should be taken before any further patients are imaged.101

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!