— The ability to mask edges of mammograms;— Low ambient light in room.As film designed for laser printing has a lower maximum optical density than does mammography film, lowerviewbox luminance is required. Therefore, a viewbox with adjustable luminance should be available.3.2.3. Quality assuranceTo ensure high quality, all of the above conditions must be met; however, in addition, it is essential that acomprehensive QA programme be in place. It is also imperative that time be allocated to allow the tests to beperformed regularly, that results be carefully recorded and that corrective action be taken promptly when indicated.The basic elements of a QA programme for digital mammography are outlined in Section 4.3.2.4. Regular maintenanceIn addition to regular QA, it is also essential that all digital mammographic units and associated equipment besubjected to regular maintenance consistent with best practice or with the recommendations of the manufacturer.19