TABLE 22. MONITOR PERFORMANCE TOLERANCESParameter Acceptable Achievable aMaximum luminance difference between paired monitors ≤10% ≤5%The contrast ratio (ratio between white target luminanceand black target luminance, including ambient light)Plot of compliance with GSDF (ΔL/L)/(jnd n – jnd n–1 )versus ((jnd n + jnd n–1 )/2)Average number of jnds per luminance intervalMaximum deviation from the average number ofjnds per luminance intervalRoot mean square deviation from the average numberof jnds per luminance intervalAmbient light level (L A ), orLuminance due to ambient light reflected fromthe monitor into radiologist’s eyes≥250:1≤10% of the response describedby DICOM soft copy GSDF≤3≤2≤1020≤ L A ≤40 lx