8.12.1.4. Interpretation of results and conclusionsTolerances:(1) There should be no significant degradation of image quality or changes in exposure factors, or in the exposureindicator for CR systems, from the baseline image.(2) The image quality assessment of the phantom should yield results that are as good as, or better than, thoseexpected with high quality screen film mammography as tested with the same phantom.8.12.1.5. Recommendations and corrective actionsIf the image quality deteriorates over time, it will be necessary to carry out other investigations (e.g. kV, AEC,display, processing algorithms) to determine the source of the change. 3636Due to the subjective component associated with the observer, it is recommended that the test always be performed by thesame person, using the same criteria and viewing conditions.135