(3) Including examinations of at least 250 patients (approximately 1000 exposures) allows reasonable statisticsfor the analysis. Collecting rejected images from a larger number of patients is encouraged because doing sowill yield more reliable data when evaluating causes for repeats.(4) Facilities that do not examine 250 patients in a quarter should still assess repeat images at least quarterly todetermine the primary causes of repeated images and reap the educational benefits of the process.(5) There is a real danger that radiographers may alter their routine procedures or criteria for accepting images ifthey know their repeated images will be analysed. This can be avoided by preserving the anonymity of theindividual radiographers.7.4.2.5. Tolerances(1) Acceptable repeat rate: