TABLE 10. MEDICAL PHYSICIST’S QUALITY CONTROL TESTS (cont.)Test Priority a Suggested frequency TolerancesEvaluation of system resolutionModulation transfer function E Annually and after detector change, etc. See Section 8.6.1.4Limiting spatial resolution E Annually and after detector change, etc. See Section 8.6.2.4X ray equipment characteristicsHalf value layer E Annually and after X ray tube change See Section 8.7.1.5Incident air kerma at the entrancesurface of PMMA slabsE Annually and after X ray tube change Not applicableDosimetryMean glandular dose (D G ) E Annually See Section 8.8.1.4Collimation systemRadiation field/image receptorcoincidenceEAnnually and after X ray tubeservice/replacementSee Section 8.9.1.4Compression paddle/breast supportalignmentEAnnually and after X ray tubeservice/replacementAcceptable: Paddle not visiblein image and edge of paddle≤5 mm beyond chest wall edgeMissing tissue at chest wall E Annually and after X ray tubeservice/replacementAchievable: ≤5 mmAcceptable: ≤7 mmImage display qualityArtefacts and uniformity (soft copy)EDAnnuallySemi-annuallySee Section 8.10.1.5Monitor luminance response andviewing conditionsViewbox luminance and viewingconditionsE Annually and after monitor service See Section 8.10.2.5, Table 22E Annually See Section 8.10.3.5Laser printer (where applicable)Artefacts and uniformityEDAnnuallySemi-annuallySee Section 8.11.1.4Film densities E Annually See Section 8.11.1.4Image qualityPhantom image quality E Annually See Section 8.12.1.4aE — essential, basic requirement; D — desirable.80

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!