FIG. 33. Image of MTF test tool mounted on a QC test object.8.6.1.4. Interpretation of results and conclusionsTolerances:(1) The spatial frequencies at which the MTF has fallen to 50 and 20% should not be less than the valuesspecified in Table 17 for the relevant model of digital mammography equipment.(2) The MTF at 2.5, 5 and 7.5 cycles/mm should not change more than 10% from the value established atcommissioning of the equipment.(3) If it is suspected that the spatial resolution may vary excessively from place to place in the image (e.g. due todetector deterioration), the MTF should be measured at different locations in the image and thesemeasurements compared.(4) A variation in noise from place to place in the image (see Section 8.5.5) can be an indicator of changes in localMTF [58].106