Chart 8 — IMAGE QUALITY WITH BREAST MIMICKING PHANTOM(This example chart may be suitable for scoring the ACR accreditation phantom)Phantom usedCR cassette size (if applicable)CR cassette ID (if applicable)Previous Image Current Image ChangeDateOperation mode (manual, semiautomatic, automatic)Target–filterkVp settingAEC detector positionDensity control (if applicable)mAsExposure index (CR)Number of fibres seenNumber of fibre-like artefactsFibres seen after deductionNumber of micro-calc. groups seenNumber of micro-like artefactsSpeck groups after deductionNumber of masses seenNumber of mass-like artefactsMasses seen after deductionTolerance*Fibres: ≥4; micro-calcifications: ≥3; masses: ≥3Pass (Y/N)?Change in fibres, microcalcifications, masses £0.5Pass (Y/N)?* The tolerances should almost certainly be more stringent for digital than for screen film mammography.Comments:Note: Use ALT+ENTER for a new line in comments box162