10.07.2015 Views

iaea human health series publications - SEDIM

iaea human health series publications - SEDIM

iaea human health series publications - SEDIM

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

8.2. MAMMOGRAPHY UNIT ASSEMBLY8.2.1. Mammography unit assembly evaluation8.2.1.1. ScopeThe mammography system contains many mechanical components. These are subject to wear or degradationover time, resulting in possible safety or performance problems. Therefore, they must be checked on a regular basis.It is also essential that the examination parameters be recorded correctly in the digital image file. For digitalmammography, an appropriately populated DICOM header complies with this labelling requirement.— Objective: To ensure that all locks, detents, angulation indicators and mechanical support devices for theX ray tube and breast support assembly are operating properly, and that the DICOM image file headers arecorrectly populated.— Frequency: Semi-annually is desirable; annually is essential.8.2.1.2. InstrumentationThermometer, preferably mounted on the wall of the digital mammography room.8.2.1.3. Methodology(1) Measure the temperature in the mammography acquisition room.(2) Verify that the freestanding dedicated mammography unit is mechanically stable under normal operatingconditions.(3) Visually inspect the unit for loose parts, cracks in the compression paddles, Bucky cleanliness and overallintegrity.(4) Check that all hoses and cables are free of breaks, crimps and knots. Hoses and cables should not be locatedunder heavy equipment.(5) Verify that all moving parts move smoothly, without undue friction; that cushions or bumpers appropriatelylimit the range of available motion; and that no obstructions hinder the full range of motion within theselimits.(6) Set and test each lock and detent independently to ensure that mechanical motion is prevented when the lockor detent is set.(7) Verify that angulation indicators function correctly.(8) Verify that the image receptor assembly is free of wobble and vibration during normal operation.(9) Verify that the CR cassette slides smoothly into the holder assembly and is held securely in any orientation(for CR operation only).(10) Verify that it is possible to override the auto-decompression function so that compression can be maintained(for procedures such as needle localizations) and that its status is displayed continuously (if autodecompressionis available).(11) Verify that compression can be manually released in the event of a power failure or automatic release failure.This verification can be done by reference to type testing data for the same model. (In the type testing, thepower supply to the equipment would be turned off with a phantom under compression and the manualcompression control would then be used to release the compression.)(12) Verify that in normal operation, the patient and operator are not exposed to sharp or rough edges or otherhazards, including electrical hazards.(13) Verify that all panel switches, indicator lights and meters are working properly.(14) Verify that the operator is protected by adequate radiation shielding during exposure.(15) Verify that current and accurate exposure technique charts are posted and confirmed by consulting with theradiographer.82

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!