(14) Examine the digital image and measure the distance from the chest wall edge of the compression plate (the ‘0’mark on the ruler or edge of coin) to the active edge of the detector using the ruler taped to the compressionpaddle.(15) Enter the results on Chart 12 in Annex II.(16) Repeat steps 2 through 15 for each image receptor size. If the unit has multiple positions for the small paddlethat result in different collimator blade positions (i.e. left and right, to allow better positioning for MLOviews), these should be evaluated.(17) Ensure that the detector is fully irradiated, that is, that the collimation does not block the edges of the activefield of the detector.(18) Repeat with the magnification table, if applicable.8.9.1.4. Interpretation of results and conclusionsTolerances:(1) Missing tissue at the chest wall:— Achievable: ≤5 mm;— Acceptable: ≤7 mm.(2) Coincidence between active detector edge and radiation field:— Achievable: The beam completely irradiates the active area of the detector but does not extend beyond thebreast support.— Acceptable: The beam completely irradiates the active area of the detector and does not extend beyond thebreast support except at the chest wall, where it can extend beyond the breast support to a maximum of 5mm. 29(3) The patient contact chest wall edge of the compression paddle should not extend beyond the chest wall edgeof the image receptor by more than 5 mm, and the chest wall edge of the paddle must not be visible in theimage.8.9.1.5. Recommendations and corrective action(1) If the missing tissue at the chest wall is greater than 7 mm or the alignment of the compression paddle and theX ray field with the active detector area does not meet the acceptable tolerances noted above, service supportis required.8.9.1.6. Time frame for corrective actionCorrective action should be taken within a period of 30 days.29Ideally, if appropriate collimation were available, the X ray field would only irradiate the entire area of the breast that lies onthe image receptor. At the time of writing such a feature is not available.118

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!