TABLE 17. ACCEPTABLE FREQUENCIES AT WHICH THE MTF FALLS TO 50% AND 20% (CYCLES/mm)(value perpendicular to chest wall/value parallel to chest wall)SystemContact mode50% 20%Agfa CR (MM3.0) 2.0/2.0 4.5/3.5Agfa CR (HM5.0) 2.5/2.0 5.5/4.5Carestream CR (EHR-M3) 2.0/2.0 4.5/4.0Carestream CR (EHR-M2) 1.5/1.5 3.5/3.0Konica CR (RP-6M) 2.5/2.0 5.0/3.5(RP-7M) 3.0/2.0 6.0/4.0(CP-1M) 3.5/2.0 7.5/4.0Fuji Amulet 4.5/4.5 7.5/4.5Fuji Profect (HR-BD) 3.0/2.0 6.0/4.0GE 2000D 2.5/2.5 5.0/5.0GE DS 3.5/3.5 6.0/6.0GE Essential 2.5/2.5 4.5/4.5Hologic Selenia 6.5/6.5 9/9IMS Giotto 4.0/4.0 6.5/6.5Philips PCREleva 5.0/5.0 9.0/8.0Planmed Nuance 4.5/5.5 9.0/8.0Sectra L30 4.0/5.5 6.0/8.0Siemens Inspiration 5.0/5.0 8.0/8.0Siemens Novation 5.0/5.0 9.0/8.08.6.1.5. Recommendations and corrective actionsIf the 50 or 20% frequencies for the MTF, either parallel or perpendicular to the chest wall, drop below thevalues given in Table 17, or if the MTF at 2.5, 5 and 7.5 cycles/mm has changed by 10% or more from previouslymeasured values, consult the manufacturer’s service representative.8.6.1.6. Time frame for corrective actionImmediately: Corrective action should be taken before any further patients are imaged.107