7.1.4. Visual inspection for artefacts (CR systems only)7.1.4.1. ScopeDust on CR plates or debris on the breast support plate or the compression plate can create artefacts thatmimic microcalcifications (Fig. 18(a)). Scratches or other defects on the plates can interfere with imageinterpretation, and dust on the optics of the plate reader can cause disturbing linear artefacts on images (Fig. 18(b)).— Objective: To ensure that clinical images are free of artefacts that might interfere with image interpretation.— Frequency: Daily, constantly throughout the day.7.1.4.2. Instrumentation(1) Cleaning cloth and solution, as recommended by the CR plate manufacturer.(2) Cleaning cloth or wipe approved for use on the breast support plate and compressor plate.7.1.4.3. Methodology(1) Inspect clinical images constantly during the day for excessive artefacts attributable to dust on the imagingplates or in the readout system, defects on the imaging plates, or dirt on the breast support plate orcompression paddle. Note that images from all plates in regular use should be inspected.(2) If artefacts are detected at a level that could interfere with the diagnostic quality of the images, inspect theplates for dust or defects.(3) Record the plate inspection and/or cleaning status on the daily and weekly checklist (Chart 1 in Annex I).(a)(b)FIG. 18. Artefacts caused by (a) dust on part of an imaging plate, and (b) scrapes (horizontal arrow) and dust in the CR readout system(vertical arrows).45

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!