QUARTERLY DIGITAL MAMMOGRAPHY REPEAT ANALYSISFacility: ABC Room: 1 Unit: Novation DRDate from: 1 Oct 2010 Date to: 31 Dec 2010Total exposures in quarter: 1201CauseNumber ofrepeat exposuresPercentage of repeats1 Positioning 12 38.72 Patient motion 8 25.83 Improper detector exposure 4 12.94 Artefact 2 6.55 X ray equipment failure 0 0.06 Software failure 1 3.27 Blank image 1 3.28 No image 0 0.09 Other 3 9.7Total # Repeat rate (%)Repeat exposures (sum of 1 through 9) 31 2.6Corrective actions: ___________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________FIG. 29. Sample repeat analysis chart. Data shown here are for illustrative purposes only.7.4.2.4. Precautions and caveats(1) All images that are repeated should be included in the repeat analysis, not just those that the radiologist askedto have repeated. Some facilities may keep repeated images in the patient study along with good images ratherthan rejecting them. These repeated images should be included in the repeat analysis.(2) At a minimum, the repeat analysis must be done at least quarterly. This process of reviewing the rejectedimages provides radiographers with educational benefits. Many facilities with higher workloads choose toconduct a repeat analysis monthly.74

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!