10.07.2015 Views

iaea human health series publications - SEDIM

iaea human health series publications - SEDIM

iaea human health series publications - SEDIM

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

8.6.2. Alternative procedure: Limiting spatial resolutionIf it is not possible to perform the MTF measurement, a less suitable alternative is to measure the limitingspatial resolution in a manner similar to that used for screen film mammography [8, 59, 60].8.6.2.1. Scope— Objective: To determine the high contrast resolution limit.— Minimum frequency: Annually and after equipment changes to the detector or X ray tube.8.6.2.2. Instrumentation(1) Star or bar resolution pattern covering at least the range 5–12 line pairs (lp)/mm.(2) PMMA slabs.(3) Magnifier lens (4× to 5× magnification).8.6.2.3. Methodology(1) Place the resolution pattern centrally on top of 45 mm of PMMA, with the closest edge 10 mm from the chestwall and with the bars oriented at approximately, but not exactly, 90° with respect to the chest wall edge of thebreast support table. The slight rotation of a few degrees is intended to avoid moiré effects.(2) Ensure that any image processing and detector correction algorithms (except for flat field correction) areturned off.(3) Confirm that the AEC sensor is not under the resolution pattern. Image the pattern using the technical factors(kV, grid, target, filter) used clinically for a compressed breast of 45 mm.(4) View the image on the monitor of the available workstation with at least a 1:1 zoom factor. If hard copy isnormally used with the system, the image could be viewed on the printed film with the aid of a magnifyinglens. Note the number of line pairs that can be observed clearly, starting with the most easily resolved.(5) Note the result on the data collection sheet.(6) Repeat the measurement with the bars oriented at approximately, but not exactly, 0°. For convenience, the twomeasurements can be done simultaneously, if the appropriate bar or star pattern is available.8.6.2.4. Interpretation of results and conclusionsTolerances:(1) Baseline value: Achievable — The limiting spatial resolution (in lp/mm) should not fall below the valueslisted under the 20% column in Table 17.(2) The limiting spatial resolution should not decrease in time by more than 10% from the baseline value.8.6.2.5. Recommendations and corrective actions(1) If variations in the resolution are observed, the cause should be identified (e.g. detector damage, change ofsize of the focal spot). If necessary, consult the manufacturer’s service representative for assistance inresolving the problem.(2) If the resolution of an image viewed on hard copy film does not meet the spatial resolution requirement,ensure that the problem is not caused by laser printing (see Section 8.11).8.6.2.6. Time frame for corrective actionImmediately: Corrective action should be taken before any further patients are imaged.108

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!