7.4.1.3. Methodology(1) Annotate the modified TG18-QC pattern image with rulers measuring the length of the horizontal and vertical5 cm rulers (arrow E, Fig. 26). This should only need to be done the first time the image is used, if theannotated image is ‘locked’ and not deleted from the workstations.(2) Print (from both the acquisition workstation and the radiologist review workstation(s)) the TG18-QC testpattern appropriate for each type of digital mammography unit used to generate images for interpretation.(3) Examine the resulting films carefully on the radiologist viewbox, using the magnifying glass.(4) Record the visibility of the different test objects on the printed image quality chart (Chart 12 in Annex I).(5) Measure the length of the 5 cm calibration lines in the horizontal and vertical directions.(6) Record the outcome of the test on the monthly, quarterly and semi-annual checklist (Chart 2 in Annex I).7.4.1.4. Interpretation of results and conclusions(1) The 0–5% contrast square and 95–100% contrast square should be distinguishable.(2) The finest horizontal and vertical line pairs should be visible in all four corners (Fig. 27).(3) The squares of different shades from black to white should be distinct.(4) Lines should appear straight and even, without apparent distortions.(5) There should be no distracting artefacts.(6) The 5 cm lines should be between 4.7 and 5.3 cm long on the printed image.7.4.1.5. Recommendations and corrective actionIf the printed film quality is unacceptable, the test should be repeated. If the results remain unacceptable, itmay be appropriate to contact the responsible medical physicist for further advice on implementing therecommendation listed below:If these results are not achieved, the printer should be serviced to correct the problem immediately, if there isno other means for primary interpretation.7.4.1.6. Time frame for corrective actionIf this test fails, do not print patient films until corrective action has been taken.FIG. 27. Detail of horizontal and vertical line pairs from TG18-QC test pattern.71

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!