TABLE 24. SPECIFICATIONS OF TEST EQUIPMENT FOR MEDICAL PHYSICIST’S TESTS (cont.)ItemCalibrated photometerDisplay test patterns (modified from AAPMTG-18)RequirementsLuminance meter (photometer) able to measure luminance between 0.5 and1000 cd/m 2 with better than 5% accuracy and a precision of at least 10 –2 , complyingwith the International Commission on Illumination (CIE) standard photopic spectralresponse to within 3%. If the monitors are LCD technology and the photometer is anear range contact design (not telescopic), the acceptance angle must be less than 5°.The photometer should not be the one attached to the graphics board on theworkstation and used to calibrate the monitors; this test is intended to be anindependent check of the system (the attached probe could be dirty or out ofcalibration). It should be auto ranging and designed to measure both luminance andilluminance. For luminance measurements, a photometric filter, lens and ambient lightshade should be incorporated. For illuminance measurements, a photometric filterwith cosine diffuser should be used.Illuminance meter able to measure illuminance between 1 and 500 lx (lm/m 2 ) withbetter than 5% accuracy, complying with the CIE standard photopic response towithin 3%.Unit specific TG18-QC comprehensive display QC pattern with DICOM header andimage format identical to that produced by the acquisition system. See the IAEA website for availability: http://<strong>human</strong><strong>health</strong>.<strong>iaea</strong>.org.Generic images of the TG18-UNL10 and TG18-UNL80 luminance uniformitypatterns d . See the IAEA web site for availability: http://<strong>human</strong><strong>health</strong>.<strong>iaea</strong>.orgNote: Regular calibration of radiation detectors is essential [66].aA smaller area of material can be used if it is raised close to the X ray tube. In this case, it must be carefully aligned parallel to theplane of the detector.bMust comply with IEC-61674 [66, 73].cIf hard copy printing is performed in the facility.dThese are uniformly grey images with brightness set at 20 and 80% of maximum brightness.142

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!