wheret 1 and t 2 are the thicknesses (in mm) of the filters used;M 0 is the average value of readings measured in steps 6 and 9 without any added filter;M 1 and M 2 are the readings measured in steps 7 and 8 that are respectively just above and just below 50%of M 0 .(2) Note the calculated HVL on Chart 10 in Annex II.8.7.1.5. Interpretation of results and conclusionsTolerance:Acceptable: kV/100 + 0.03 ≤ HVL ≤ kV/100 + CwhereC = 0.12 for Mo/Mo;0.19 for Mo/Rh;0.22 for Rh/Rh;0.30 for W/Rh;0.32 for W/Ag;0.25 for W/Al;and kV is the measured value for the nominal kV selected.On Chart 10 in Annex II, record whether the value of the HVL is acceptable. If this tolerance is met, it isgenerally the case that regulatory requirements for total beam filtration without the compression paddle willalso be satisfied.8.7.1.6. Recommendations and corrective actionsIf the HVL is very low or very high, investigative action should be undertaken. This may include ameasurement of the kV to confirm that the tube potential is properly calibrated. Methods are described in Ref. [34].8.7.1.7. Time frame for corrective actionImmediately: Corrective action should be taken until the problem is solved.110