7.4. QUARTERLY TESTS7.4.1. Printed image quality7.4.1.1. Scope— Objective: To ensure that the printed image quality is consistently high.— Frequency: Quarterly and when reduced printed image quality is suspected. This test applies only to sitesprinting hard copies.7.4.1.2. Instrumentation(1) Magnifying lens (4× to 5× magnification).(2) Ruler.(3) Modified TG18-QC patterns (Fig. 26). Use test patterns that are specific to the digital mammography units inthe facility.FIG. 26. Modified TG18-QC image with test objects indicated: A — 0–5% contrast square; B — 95–100% contrast square;C — horizontal and vertical line pairs; D — squares going from black to white; E — 5 cm lines.70