(6) Measure the luminance at each point. Record the values on the data collection sheet.(7) Repeat the tests for all viewboxes used for interpreting printed mammograms and for the viewboxes used bythe radiographer to check the printed mammograms during the examination.8.10.3.4. Calculation procedureFor a single viewbox or single panel in a viewbox bank, select the central luminance value (L c ) and the mostdiscrepant reading recorded (L disc ). Apply the following expression:Maximum deviation (%) = 100 LdiscL- Lcc(10)For viewboxes consisting of several panels or for viewboxes adjacent to one another, the maximum deviationbetween the central luminance value of any panel, L cx , and the calculated mean of the central luminance of allpanels, L mean , is found by applying the following expression:Maximum deviation (%) = 100 Lcx- LLmeanmean(11)8.10.3.5. Interpretation of results and conclusions(1) Luminance: Acceptable — maximum luminance for each panel: greater than 3000 cd/m 2 (nit). 32(2) Luminance uniformity: Acceptable — less than 30% maximum deviation for different areas of a singleviewbox or single panel in a viewbox bank, and less than 15% maximum deviation between central luminanceof panels in a viewbox bank or between adjacent viewboxes.(3) Room illuminance: Acceptable — 20–40 lx.(4) A convenient and effective method for masking viewboxes to exclude bright light around the edges ofmammograms should be available.8.10.3.6. Recommendations and corrective actions(1) If the ambient illumination conditions are greater than the recommended values, contact the person in chargeof the maintenance service to modify the illumination of the area.(2) If the luminance level of the viewbox is less than 3000 cd/m 2 , or if the luminance or the colour of light of anindividual lamp appears significantly different from those of others in the same viewbox, all lamps in theviewbox should be replaced.Note: Tri phosphor or quad phosphor tubes are recommended because of their increased luminance output. Ifpossible, all tubes in a viewbox or viewbox bank should be replaced at the same time to ensureuniformity. It is advisable to purchase fluorescent tubes in one batch to ensure that the colour matchesbetween the tubes.32If the viewbox is to be used only for viewing laser printed hard copy film, a luminance of less than 3000 cd/m 2 is probablyacceptable.130

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!