6.5 Setting up for the experiments<br />

In the previous chapter, the capabilites and set-up <strong>of</strong> the analysers was described. In this<br />

section, I will outline how the analysers were calibrated prior to each subject in the<br />

methodological experiments and the basic procedure to serve as an introduction to all the<br />

subsequent experiments on adult volunteer subjects. Changes in this exact procedure and<br />

results <strong>of</strong> each set <strong>of</strong> studies will be outlined as separate experiments. I will present and then<br />

discuss the initial findings after the first investigations which compared the two sampling<br />

techniques with NO, CO2, expiratory mouth pressure, and (in the t-piece system) expiratory<br />

flow results.<br />

6.5.1 Set-up and calibrations<br />

<strong>The</strong>se instructions were set out by Carolyn Busst and myself after many individual tests <strong>of</strong> the<br />

machinery and then posted next to the NO analyser to follow each time I did these studies to<br />

ensure a standard procedure. I have re-presented them here as they were written in the present<br />

tense and with numerical representation <strong>of</strong> the recorded numbers for times and frequency or<br />

repetitions, but have added in italics some additional notes that came to be part <strong>of</strong> the set-up<br />

as I progressed through and became more experienced. <strong>The</strong>re was no change to the nature <strong>of</strong><br />

the investigations, but made it easier to standardise the machines and have the set-up running<br />

efficiently. <strong>The</strong> purpose <strong>of</strong> these procedures was two-fold; to calibrate the analysers, in<br />

particular the NO and COz analysers, and to mark the levels on the chart recorder to enable<br />

calculation <strong>of</strong> the study parameters.<br />

One hour before the study:<br />

1. Check that the NO analyser and the vacuum pump are switched on and that the ozone<br />

generator in the analyser is switched on (toggle switch left hand side, front panel). [NB: I<br />

found the analyser and vacuutn more stable and took less time to reach a stable level if I<br />

let it run continuously Monday to Friday or over weekends as well if experiments were<br />

being underraken. This is in keeping with how it would have been used oiginally as a<br />

continuous monitor <strong>of</strong> air NO pollution.l<br />

2. Switch on the capnograph (switch located at the back <strong>of</strong> the machine).<br />

g. Switch on the electrospirometer (switch located at the back <strong>of</strong> the machine).<br />

4. Check that the calibration gas cylinders and air cylinders are not empty'<br />

S. Check the paper in the chart recorder. If in doubt regarding having sufficient to record the<br />

whole next subject experiment, replace with a new roll. Keep the old ones for test runs.<br />

r3l