6. Check that the pens work. Fill with ink if necessary, careful not to overfill as this can lead<br />

to smudges on the traces.<br />

7. Check that the mouthpiece is clean and that the sample tubing is free from water. If it is<br />

not, flush the tubing with compressed air from the cylinder.<br />

8. Do not switch on the pressure equipment as it is battery powered and this will just waste<br />

the battery.<br />

Calibration just before the study:<br />

On chart recorder note:<br />

o Date.<br />

o Study name.<br />

o Subject name.<br />

o Parameters and voltage scales [these remained unchanged and would only have mattered<br />

if there was a log scale change in the concentrations <strong>of</strong> the gases, pressure orflow that we<br />

were measuring, see NO calibration belowl.<br />

o Paper speed.<br />

o Barometric pressure.<br />

o Temperature.<br />

o <strong>The</strong> background ambient NO level.<br />

<strong>The</strong> procedures for calibrating the analysers:<br />

Calibration <strong>of</strong> the NO analyser:<br />

o Make sure the rotameter valve is closed.<br />

o Tnro calibration: connect NO sampling probe to the black, cylindrical NO charcoal scrub<br />

unit.<br />

o NO concentration calibration one: connect NO sampling probe to the calibration gas 'one'<br />

nozzle - turn on the cylinder and record the level noting on the chart record.<br />

o NO concentration calibration two: connect NO sampling probe to the calibration gas 'two'<br />

nozzle - turn on the cylinder and record the level noting on the chart record.<br />

o Turn <strong>of</strong>f the cylinders.<br />

o Connect the sample port to the mouth-piece.<br />

o Run the chart recorder to record both NO and COz baseline ambienUroom air values and<br />

note results on the chart record.<br />

132