05.06.2013 Views

PNNL-13501 - Pacific Northwest National Laboratory

PNNL-13501 - Pacific Northwest National Laboratory

PNNL-13501 - Pacific Northwest National Laboratory

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

concentration of 0.5% AMX was the most effective. In<br />

Figure 1, results are shown for an injection experiment<br />

conducted in a 1.0-meter long column of Hanford<br />

sediment. This experiment was conducted by first<br />

pretreating the column with approximately two pore<br />

volumes of a solution containing only 0.5% AMX at a pH<br />

of 8.0. The injection slurry contained 0.5% AMX and<br />

1.0% Fe 0 at pH 8.0. Three pore volumes of this slurry<br />

were injected at 7.2 cm/min. The results indicate a fairly<br />

even distribution of iron throughout the column, ranging<br />

from 0.5% Fe 0 at 10 cm to about Fe 0 0.3% at the effluent<br />

end of the column.<br />

% Fe 0<br />

1.0<br />

0.8<br />

0.6<br />

0.4<br />

0.2<br />

0.0<br />

0 20 40 60 80 100<br />

Distance (cm)<br />

Figure 1. Distribution of Fe 0 in a 100-cm column of Hanford<br />

sediment resulting from injection of 3 pore volumes of slurry<br />

containing 1.0 % Fe 0 , and 0.5 % AMX, using a flow rate of<br />

7.2 cm/min, and a 2-pore volume pretreatment with 0.5 %<br />

AMX<br />

The results for an experiment in which the pretreatment<br />

step was skipped are shown in Figure 2. In this case, the<br />

Fe 0 distribution is much less even, with lower<br />

concentrations observed in the first part of the column.<br />

Another experiment in which the injection rate was<br />

reduced to 1.6 cm/min is shown in Figure 3. In this case<br />

we can see that the iron concentrations were significantly<br />

higher than for the higher flow rate experiment illustrated<br />

in Figure 1. In addition, the Fe 0 concentrations were<br />

observed to drop off rapidly as a function of distance from<br />

the inlet.<br />

To determine if the injection of Fe 0 into the columns<br />

caused any permanent reduction in permeability, a<br />

column injection experiment was conducted in an<br />

identical fashion to that of the experiment illustrated in<br />

Figure 1. However, in this case the inlet pressure was<br />

monitored prior to, during, and after the injection phase of<br />

the experiment. Prior to and after the pretreatment and<br />

Fe 0 slurry injection phase of the experiment, water was<br />

pumped through at a much slower flow rate. As can be<br />

256 FY 2000 <strong>Laboratory</strong> Directed Research and Development Annual Report<br />

% Fe 0<br />

1.0<br />

0.8<br />

0.6<br />

0.4<br />

0.2<br />

0.0<br />

0 20 40 60 80 100<br />

Distance (cm)<br />

Figure 2. Distribution of Fe 0 in a 100-cm column of Hanford<br />

sediment resulting from injection of 3 pore volumes of slurry<br />

containing 1.0 % Fe 0 , and 0.5 % AMX, using a flow rate of<br />

7.4 cm/min, and no pretreatment with 0.5 % AMX<br />

% Fe 0<br />

4<br />

2<br />

0<br />

0 20 40 60 80 100<br />

Distance (cm)<br />

Figure 3. Distribution of Fe 0 in a 100-cm column of Hanford<br />

sediment resulting from injection of 3 pore volumes of slurry<br />

containing 1.0 % Fe 0 , and 0.5 % AMX, using a flow rate of<br />

1.6 cm/min, and a 2-pore volume pretreatment with 0.5 %<br />

AMX<br />

seen from Figure 4, injection of water at 0.76 cm/min<br />

resulted in a very low back pressure at the column inlet<br />

(0.7 psig). As the 0.5% AMX pretreatment solution was<br />

injected into the column at a flow rate of 7.0 cm/min, the<br />

back pressure began to increase. During the slurry<br />

injection, the back pressure continues to increase. After<br />

the injection phase of the experiment, water at a flow rate<br />

of 0.85 cm/min was pumped through the column. When<br />

this was done, the back pressure decreased significantly<br />

from about 70 psig to 11 psig. This is a significant<br />

decrease in back pressure; however, it is significantly<br />

higher than the initial back pressure measured prior to the<br />

injection phase. After 3.4 pore volumes of water were<br />

pumped through the column, the back pressure returned to

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!