ASXPRESS system. To do so, verify proper operation of all manual functions by<br />
completing the following steps:<br />
1 Operate manual features of the Xpress Config Tool software.<br />
With the ASXPRESS system connected to the autosampler/ICP/ICP-MS systems,<br />
the autosampler “queried” (Chapters 3 and 4), and the manual functions<br />
showing in the Xpress Config software window:<br />
o<br />
o<br />
o<br />
o<br />
o<br />
o<br />
o<br />
Click the “<strong>Manual</strong> Cycle” button. The ASXPRESS system should run through a<br />
complete cycle (“Load” and “Inject”), operating the autosampler peristaltic<br />
pump and the ASXPRESS components (6-Port Valve and Vacuum Pump).<br />
Click the “Load” button. The ASXPRESS system should move the 6-Port Valve<br />
to the “Load” position (no other apparatus operates).<br />
Click the “Inject” button. The ASXPRESS system should move the 6-Port<br />
Valve to the “Inject” position (no other apparatus operates).<br />
Click the “On” (autosampler pump control) button. The autosampler<br />
peristaltic rinse pump should turn on (no other apparatus operates).<br />
Click the “Off” (autosampler pump control) button. The autosampler<br />
peristaltic rinse pump should turn off (no other apparatus operates).<br />
Click the “On” (ASXPRESS vacuum pump control) button. The ASXPRESS<br />
vacuum pump should turn on (no other apparatus operates).<br />
Click the “On” (ASXPRESS vacuum pump control) button. The ASXPRESS<br />
vacuum pump should turn on (no other apparatus operates).<br />
Figure 1-54<br />
functions<br />
Xpress Config Tool—Expanded to access manual<br />
107