<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />
Chapter 5: Using the Rapid Sample Introduction System<br />
• Observe the same general electrostatic discharge precautions as with<br />
any other integrated circuit electronic devices. Low humidity<br />
environments, especially when combined with static-generating<br />
materials, require the use of maximum care.<br />
WARNING<br />
Discharge static buildup and ground to the ASXPRESS Rapid Sample<br />
Introduction System base or cabinet before performing any<br />
maintenance. Do not touch or short-circuit bare contacts.<br />
Avoid using the ASXPRESS Rapid Sample Introduction System if strong<br />
electromagnetic interference, radio frequency interference, or radioactivity is<br />
present. Interference fields can cause erratic. The ASXPRESS Rapid Sample<br />
Introduction System will not function properly if the level of radioactivity is<br />
above background.<br />
5.1.2 Replacing Rapid Sample Introduction System<br />
Components<br />
The following ASXPRESS Rapid Sample Introduction System components wear<br />
out under normal use and must be replaced periodically. These components<br />
are consumables and replacement is not covered under warranty. The<br />
lifecycle of these components vary considerably depending upon the<br />
application and solids content of the sample being transported. Approximate<br />
lifecycle ranges are listed as follows:<br />
‣ 6-Port Injection Valve (5K – 60K cycles – dependant on solids levels)<br />
‣ Vacuum Pump (10k – 500K cycles – dependant on solids levels)<br />
‣ Tubing (user/application determined)<br />
‣ Sample Loop (user/application determined)<br />
If you fail to replace these components when/before they deteriorate, the system<br />
will not function properly. For information about replacing components, see<br />
Chapter 6, “Maintaining the Rapid Sample Introduction System.”<br />
5.1.3 Purchasing Supplies<br />
When you need to purchase additional supplies, it is extremely important that<br />
you choose the appropriate sizes and materials.<br />
To order additional supplies, refer to the <strong>CETAC</strong> Accessories and Supplies<br />
Catalog for the ASXPRESS Rapid Sample Introduction System.<br />