ASXpress Operator's Manual - CETAC Technologies
ASXpress Operator's Manual - CETAC Technologies
ASXpress Operator's Manual - CETAC Technologies
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />
Chapter 3: Installing the Rapid Sample Introduction System<br />
Figure 1-47<br />
Rinse Tubing Diagram<br />
To retrofit the rinse station plumbing on an ASX-520 or ASX-260, complete the<br />
following:<br />
1. Gently pull the existing rinse station away from the autosampler rinse<br />
block.<br />
2. Remove the rinse tubing between the rinse station and the peristaltic<br />
pump and also remove the rinse solution suction tubing to the<br />
peristaltic pump.<br />
3. Install the Y-tubing assembly (supplied in the upgrade kit) on the<br />
existing Standard Bowl Rinse Station, and onto the lower two<br />
connections of the peristaltic pump, as shown in Figure 3-43 and 3-44.<br />
4. Connect one end of the supplied drain tubing to the drain/overflow<br />
fitting (highest fitting) of the rinse station. Connect the drain tube from<br />
the rinse station to a suitable waste container, ensuring that the end of<br />
the drain tube does not become submerged below the waste liquid<br />
level, as that condition will impede proper drainage. Note: This length<br />
54