<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />

Chapter 3: Installing the Rapid Sample Introduction System<br />

Figure 1-47<br />

Rinse Tubing Diagram<br />

To retrofit the rinse station plumbing on an ASX-520 or ASX-260, complete the<br />

following:<br />

1. Gently pull the existing rinse station away from the autosampler rinse<br />

block.<br />

2. Remove the rinse tubing between the rinse station and the peristaltic<br />

pump and also remove the rinse solution suction tubing to the<br />

peristaltic pump.<br />

3. Install the Y-tubing assembly (supplied in the upgrade kit) on the<br />

existing Standard Bowl Rinse Station, and onto the lower two<br />

connections of the peristaltic pump, as shown in Figure 3-43 and 3-44.<br />

4. Connect one end of the supplied drain tubing to the drain/overflow<br />

fitting (highest fitting) of the rinse station. Connect the drain tube from<br />

the rinse station to a suitable waste container, ensuring that the end of<br />

the drain tube does not become submerged below the waste liquid<br />

level, as that condition will impede proper drainage. Note: This length<br />

54