ASXpress Operator's Manual - CETAC Technologies
ASXpress Operator's Manual - CETAC Technologies
ASXpress Operator's Manual - CETAC Technologies
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />
Chapter 9: Guide to Operating a <strong>CETAC</strong> Autosampler Using a Terminal<br />
Program<br />
10. Turn on the autosampler.<br />
The HyperTerminal window should display an OK.<br />
9.3 Autosampler Commands<br />
The following commands will produce various responses of the autosampler.<br />
Command<br />
Ver<br />
Home<br />
Tray=n<br />
Tube=row-col-down<br />
Pmp on<br />
Pmp off<br />
Rinse<br />
Down=n<br />
Up<br />
Description<br />
Returns firmware version.<br />
Returns all axis to home position, same<br />
as power up.<br />
Defines tray size and n= #of positions.<br />
Legal values are 21, 24, 40, 60, and 90.<br />
Tube=row-column-down as defined by<br />
tray command. Example: Tube=3-4-<br />
150<br />
Pump on if unit has a pump.<br />
Pump off if unit has a pump.<br />
Moves sipper to the rinse position,<br />
extends and retracts the sipper 3 times<br />
and starts rinse pump. Stays in down<br />
position with pump running. up,pmp<br />
off stops the pump.<br />
Moves the z-axis down by the<br />
parameter(n) in mm. Do not run Down<br />
command if sipper is not all the way up<br />
on up position or damage may occur to<br />
sipper or z-axis.<br />
Moves z-axis to upper most postion.<br />
Table 9-1<br />
Autosampler Commands<br />
You can use these commands to determine if the <strong>CETAC</strong> autosampler is<br />
communicating and functioning properly. If more assistance is needed, please<br />
contact <strong>CETAC</strong> customer service.<br />
123