Chapter 3: Installing the Rapid Sample Introduction System<br />

Figure 1-49<br />

ASXPRESS Valve/Pump Module Tubing Connections<br />

To make the plumbing connections to the ASXPRESS (Figure<br />

1-49), perform the following:<br />

1. Remove the protective cover from the 6-port valve.<br />

2. Install a sample loop between ports #1 and #4 of the 6-port valve (as<br />

labeled on the valve). Use caution to prevent kinking of loop!<br />

3. Ensure that the 60” length of 1/8” tubing (which is pre-connected to the<br />

“out” port of the vacuum pump) is inserted into a waste container. Cut<br />

this length to fit your specific setup, ensuring that the discharge end is<br />

not placed so that it may become submerged below the liquid level of the<br />

waste container.<br />

4. Install the autosampler probe in the autosampler per instructions<br />

included with the autosampler.<br />

5. Connect the tube from the probe to port #2 of the 6-port valve.<br />