<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />
Chapter 3: Installing the Rapid Sample Introduction System<br />
3. On the ASX-260 board, move jumper JP2 to JP4 and leave JP3 in<br />
place (Figure 3-61 shows jumpers).<br />
Figure 1-61<br />
Jumpers on board in normal position (ASX-260)<br />
3.4.2 Load the Firmware Upgrade, via the Rabbit Field<br />
Utility (RFU) program, from the CD provided.<br />
The software required to upgrade/load the firmware on the autosampler and<br />
the firmware upgrade files are located on the installation CD provided with the<br />
ASXPRESS system. The overview of the following sections is as follows:<br />
• Section 3.4.2.1 covers the process required to ensure that proper<br />
communication exists between the autosampler and host computer.<br />
• Section 3.4.2.2 covers the process for setting up the Rabbit Field Utility<br />
and for choosing the proper file to be loaded to the rabbit module.<br />
• Section 3.4.2.3 covers the process for loading (flashing) the selected<br />
firmware file to the rabbit module.<br />
• Section 3.4.2.4 covers the process for returning the autosampler to<br />
normal operation.<br />
• Section 3.4.2.5 covers verification of the firmware upgrade.<br />
3.4.2.1 Establish Communications<br />
1. Ensure that the Z-drive assembly is properly installed on the<br />
autosampler Y-arm.<br />
2. Connect the power cord to the autosampler power supply, connect<br />
the serial cable to the host computer and turn the autosampler on.<br />
3. Start HyperTerminal. The path for starting HyperTerminal on a<br />
Windows XP equipped PC are as follows:<br />
72