<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />
Chapter 6: Maintaining the Rapid Sample Introduction System<br />
Note:<br />
It is recommended to clean the 6-port valve every 20,000 cycles.<br />
However, the frequency of cleaning interval will vary depending on<br />
application.<br />
Materials required:<br />
• dust-free work surface<br />
• hex key wrench (provided)<br />
• lint-free clean room wipes<br />
• low-pressure compressed clean air (such as a canned air duster<br />
product)<br />
To clean the valve, complete the following steps:<br />
1 Unplug the ASXPRESS Rapid Sample Introduction System.<br />
2 Remove all tubing and connectors from the 6-port valve.<br />
3 Use the hex key wrench provided to remove the three screws on the<br />
front of the 6-port valve<br />
Do not remove the valve body from the actuator as it will lose its position and<br />
will require retraining if removed. Valve retraining will be discussed in detail<br />
on page Error! Bookmark not defined..<br />
4 Note the orientation of valve body components prior to removal.<br />
5 Gently remove the stator of the valve.<br />
6 Use a clean-room wiper to gently clean the channels and surfaces of<br />
the stator and rotor of any debris.<br />
7 Use low-pressure, canned clean air to blow the channels and ports free<br />
of any remaining debris.<br />
8 Very carefully replace the stator of the valve, in the proper orientation<br />
noted in step 4, and tighten the three screws to set firmly in place.<br />
9 Reinstall all tubing and check for leaks using the Xpress Config Tool to<br />
“<strong>Manual</strong> Cycle” the ASXPRESS system as described in Section 4 of this<br />
manual.<br />
97