<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />
Chapter 3: Installing the Rapid Sample Introduction System<br />
3.1.6 Install the Standard-Bowl Rinse Station and Modified<br />
Rinse Tubing Arrangement<br />
A Standard-Bowl Rinse Station (Figure 1-45) and associated modified tubing<br />
arrangement is required to accommodate the design requirements of the<br />
ASXPRESS system. Please note the following considerations:<br />
‣ If you have purchased a full system, it will already be configured<br />
properly and this step will not be necessary.<br />
‣ If you are connecting the ASXPRESS system to an existing ASX-260 or<br />
ASX-520 (standard speed) autosampler, the proper rinse station will<br />
already be in place, but tubing modifications (outlined in this section)<br />
will be required.<br />
‣ If you are connecting the ASXPRESS system to an ASX-520HS ( high<br />
speed autosampler), it will have the large-bowl rinse station installed,<br />
and you will need to replace it with the standard-bowl rinse station<br />
(supplied with the upgrade kit), as well as to make the tubing<br />
modifications (Figure 1-46 and Figure 1-47) as are outlined in this<br />
section.<br />
Figure 1-45<br />
Standard-Bowl Rinse Station<br />
52