<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />

Chapter 10: Safety and Regulatory Information<br />

10 Safety and Regulatory Information<br />

Review this product and related documentation to familiarize with safety<br />

markings and instructions before you operate the instrument.<br />

10.1 Characteristics<br />

10.1.1 Environmental Characteristics<br />

Operating Temperature +13° C to +30° C (+55° F to +85° F)<br />

Non-Operating Temperature +0° C to +55° C (+32° to +131° F)<br />

Operating Altitude<br />

Relative Humidity<br />

Up to 2,000 m (6,562 ft)<br />

0% to 60% non-condensing<br />

Pollution Degree Pollution Degree 2<br />

Table 9-2: Environmental Characteristics<br />

For indoor use only.<br />

Normally no pollution or only dry, nonconductive<br />

pollution occurs. The<br />

pollution has no influence. Occasionally,<br />

however, a temporary conductivity<br />

caused by condensation may be<br />

expected.<br />

Avoid sudden, extreme temperature changes which could cause condensation<br />

on circuit boards in the product.<br />

See page 10 for information on chemical compatibility.<br />

10.1.2 Electrical Characteristics<br />

10.1.2.1 Power requirements<br />

Electronics Module<br />

Input:<br />

Power is provided by the <strong>CETAC</strong> autosampler through the<br />

power/signal cable which is attached to the electronics<br />

module.<br />

DC Voltage and Current<br />

24 V<br />

2.5 A<br />

Installation Category: CAT I (Mains isolated)<br />

Output:<br />

Connector for the valve/pump module.<br />

Valve/Pump Module<br />

Input:<br />

Use only with the provided electronics module.<br />

Installation Category: CAT I (Mains isolated)<br />

Table 9-3: Power Requirements<br />