Chapter 4: Verifying Installation/ Function Overview<br />
timing parameters.<br />
7 Click on the “Inject” button (Figure 4-3); doing this so will position the<br />
ASXPRESS 6-port valve into the mode in which sample injection into the<br />
ICP takes place (via the Peristaltic pump on the ICP), and will<br />
illuminate the “Inject” Led on the front panel of the ASXPRESS<br />
electronics module (no autosampler operation takes place at this<br />
time).<br />
8 Click the “<strong>Manual</strong> Cycle” button (Figures 4-2 and 4-3); doing this will<br />
perform one complete cycle of system operation, operating each<br />
component on the ASXPRESS system and autosampler, using the saved<br />
or queried parameters of the Configuration Tool software settings<br />
(See “Principal of Operation” located on the CD provided with the<br />
ASXPRESS system). This is also a useful function for loading the tubing<br />
and priming the rinse pump.<br />
9 Click on the other manual function radio buttons (autosampler pump<br />
control and vacuum pump control), (Figure 4-3) cycling them on and<br />
off to verify proper operation.<br />
Figure 1-51 Xpress Configuration Tool—“Query <strong>ASXpress</strong><br />
Params” (post-query)<br />