<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />

Chapter 3: Installing the Rapid Sample Introduction System<br />

6. Connect the 14” length of 0.030” x 0.062” tubing (which is pre-installed<br />

to the green connector at port #6 of the 6-port valve) to the ICP / ICP-MS<br />

peristaltic pump. This length of tubing is fitted with a tee on its end<br />

which is intended to be connected to two individual output (discharge)<br />

channels of the ICP / ICP-MS peristaltic pump. The input (suction) ports<br />

of this peristaltic pump are connected as follows:<br />

• One channel to carrier/rinse solution<br />

• The other channel open to air<br />

This plumbing configuration is crucial to the proper operation of the ASXPRESS<br />

system and must be followed.<br />

7. Connect the 14” length of 0.030” x 0.062” tubing (which is pre-installed<br />

to the blue connector at port #5 of the 6-port valve) to the nebulizer.<br />

Place the ASXPRESS valve/pump module as close to the nebulizer as is<br />

possible, and cut the length of tubing to the shortest length possible<br />

which will allow connection of port #5 to the nebulizer without any<br />

kinking. The shortest length tubing at this connection is most desirable,<br />

as it provides the greatest time savings while operating the ASXPRESS<br />

system.<br />

Note:<br />

The ASXPRESS system is designed to operate and interface with many<br />

manufacturers’ instruments. To allow for the closest proximity<br />

placement of the ASXPRESS valve/pump module to the instruments<br />

nebulizer, <strong>CETAC</strong> <strong>Technologies</strong> offers a modular articulating support<br />

stand, which allows for stable mounting and flexible positioning of the<br />

valve/pump module. Contact <strong>CETAC</strong> <strong>Technologies</strong> for technical and<br />

ordering information.<br />

59