Chapter 3: Installing the Rapid Sample Introduction System<br />
6. Connect the 14” length of 0.030” x 0.062” tubing (which is pre-installed<br />
to the green connector at port #6 of the 6-port valve) to the ICP / ICP-MS<br />
peristaltic pump. This length of tubing is fitted with a tee on its end<br />
which is intended to be connected to two individual output (discharge)<br />
channels of the ICP / ICP-MS peristaltic pump. The input (suction) ports<br />
of this peristaltic pump are connected as follows:<br />
• One channel to carrier/rinse solution<br />
• The other channel open to air<br />
This plumbing configuration is crucial to the proper operation of the ASXPRESS<br />
system and must be followed.<br />
7. Connect the 14” length of 0.030” x 0.062” tubing (which is pre-installed<br />
to the blue connector at port #5 of the 6-port valve) to the nebulizer.<br />
Place the ASXPRESS valve/pump module as close to the nebulizer as is<br />
possible, and cut the length of tubing to the shortest length possible<br />
which will allow connection of port #5 to the nebulizer without any<br />
kinking. The shortest length tubing at this connection is most desirable,<br />
as it provides the greatest time savings while operating the ASXPRESS<br />
system.<br />
Note:<br />
The ASXPRESS system is designed to operate and interface with many<br />
manufacturers’ instruments. To allow for the closest proximity<br />
placement of the ASXPRESS valve/pump module to the instruments<br />
nebulizer, <strong>CETAC</strong> <strong>Technologies</strong> offers a modular articulating support<br />
stand, which allows for stable mounting and flexible positioning of the<br />
valve/pump module. Contact <strong>CETAC</strong> <strong>Technologies</strong> for technical and<br />
ordering information.<br />