<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />

Chapter 3: Installing the Rapid Sample Introduction System<br />

Figure 1-42<br />

Mainboard shown with rabbit module removed<br />

Proceed to the “Increase the Peristaltic Pump Speed” section of this manual<br />

(Do not replace covers, etc., at this time).<br />

Note:<br />

Do not replace the cover/shields/Z-drive assembly on the autosampler<br />

until the following pump speed increase operation is also completed.<br />

3.1.4 Increase the Peristaltic Pump Speed Setting<br />

In order to provide the necessary additional flow rate at the rinse station, the<br />

pump speed must be increased. This is achieved by changing the position of a<br />

jumper on the autosampler main board.<br />

The jumper must be moved from its existing position to the position marked<br />

“Max RPM” (Figure 1-43), on the ASX-520/ASX-520HS board, and to the top<br />

position (see note below) of the bank of jumper pins marked J6 (Figure 1-44)<br />

on the ASX-260 board. This jumper/pin arrangement is located on the left side<br />

of the ASX-520/ASX-520HS board, and near the center of the ASX-260 board, as<br />

shown in Figure 3-41.<br />

49