Chapter 3: Installing the Rapid Sample Introduction System<br />
Figure 1-42<br />
Mainboard shown with rabbit module removed<br />
Proceed to the “Increase the Peristaltic Pump Speed” section of this manual<br />
(Do not replace covers, etc., at this time).<br />
Note:<br />
Do not replace the cover/shields/Z-drive assembly on the autosampler<br />
until the following pump speed increase operation is also completed.<br />
3.1.4 Increase the Peristaltic Pump Speed Setting<br />
In order to provide the necessary additional flow rate at the rinse station, the<br />
pump speed must be increased. This is achieved by changing the position of a<br />
jumper on the autosampler main board.<br />
The jumper must be moved from its existing position to the position marked<br />
“Max RPM” (Figure 1-43), on the ASX-520/ASX-520HS board, and to the top<br />
position (see note below) of the bank of jumper pins marked J6 (Figure 1-44)<br />
on the ASX-260 board. This jumper/pin arrangement is located on the left side<br />
of the ASX-520/ASX-520HS board, and near the center of the ASX-260 board, as<br />
shown in Figure 3-41.<br />
49