<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />

Chapter 5: Using the Rapid Sample Introduction System<br />

5.2 Shutting Down the ASXPRESS Rapid Sample<br />

Introduction System<br />

To shut down the ASXPRESS Rapid Sample Introduction System, complete the<br />

following steps:<br />

1. Drain the system by removing the rinse solution uptake tubing from the<br />

rinse solution sources (autosampler and ICP/ICP-MS). Operate the<br />

autosampler peristaltic pump and ASXPRESS vacuum pump by operating<br />

the “manual cycle” function or by using the manual function for<br />

operating the autosampler peristaltic pump (both from within the<br />

Xpress Configuration Tool software) and manually running the ICP/ICP-<br />

MS peristaltic pump until all solution drains from the tube attached to<br />

the rinse station outlet, the vacuum pump waste discharge, and from<br />

the ICP/ICP-MS nebulizer drain.<br />

Note:<br />

If using a rinse solution other than deionized water, flush the rinse<br />

system with deionized water before shutting down and/or draining the<br />

ASXPRESS discrete Sampling Accessory. For more information, see the<br />

following section, “Flushing the Rinse Station and Flow Path”.<br />

2. Switch the power switch on the back of the autosampler to “Off.”<br />

93