Chapter 4: Verifying Installation/ Function Overview<br />
4.1.1 To test the communications interface, complete the<br />
following steps:<br />
Test the operation of the ASXPRESS system, and its interface with the<br />
Autosampler, by switching modes via the Configuration Tool.<br />
1 Open the Xpress Configuration Tool Software (Figure 1-50).<br />
Xpress Configuration Tool Software Window (pre-<br />
Figure 1-50<br />
query)<br />
2 Click on the “Query <strong>ASXpress</strong> Params” button (this will communicate<br />
with the autosampler/<strong>ASXpress</strong>, returning default values in selection<br />
boxes, and will make the “<strong>Manual</strong> Cycle” and “double arrow” arrow<br />
buttons available for selection (Figure 1-50).<br />
3 Check the “Enable <strong>ASXpress</strong> Operation” box (Figure 1-51). Unchecking<br />
this box would allow the autosampler to run independently of the<br />
<strong>ASXpress</strong> system, as if the autosampler were connected directly to the<br />
host PC. (Note that a different tubing arrangement would be required<br />
to run the autosampler without the <strong>ASXpress</strong> system.)<br />
4 Click on the “double arrow” button at bottom right of the window,<br />
expanding it to reveal the manual commands (Figure 1-52).<br />
5 Click on the “Load” button (Figure 4-3); doing so will position the<br />
ASXPRESS 6-port valve into the mode which loads the sample loop, and<br />
will illuminate the “Load” Led on the front panel of the <strong>ASXpress</strong><br />
electronics module (no autosampler operation takes place at this<br />
time).<br />
6 Click the “Save to ASX” button once. This will load an initial set of<br />
83