<strong>ASXpress</strong> Rapid Sample Introduction System Operator’s <strong>Manual</strong><br />

Chapter 5: Using the Rapid Sample Introduction System<br />

5 Using the Rapid Sample Introduction<br />

System<br />

The ASXPRESS Rapid Sample Introduction System is both reliable and easy to<br />

use. Before using it, however, ensure that your lab environment provides<br />

operating conditions that will prolong the life of the ASXPRESS system. Once the<br />

proper operating conditions are met, you can operate the rapid sample<br />

introduction system. When you finish using the ASXPRESS system, it may be<br />

necessary to flush the rinse station and flow path before shutting the system<br />

down.<br />

This chapter explains how to create the proper operating conditions for using<br />

the ASXPRESS Rapid Sample Introduction System. It also explains how to start<br />

and shut down the system, and how to flush the rinse station and flow path.<br />

5.1 Establishing Optimal Operating Conditions<br />

The ASXPRESS Rapid Sample Introduction System operates reliably under<br />

virtually the same operating conditions described in the autosampler<br />

Operator’s <strong>Manual</strong>. However, it is not indestructible. Malfunction or damage<br />

can occur if specific operating conditions are not met. Meeting these<br />

conditions requires that you create the proper lab environment, replace<br />

ASXPRESS Rapid Sample Introduction System components that wear out under<br />

normal use, and purchase the appropriate supplies. The following sections<br />

explain how to meet these conditions.<br />

Note:<br />

Damage or malfunction that results from exposure to inappropriate<br />

operating conditions may constitute misuse and abuse and be excluded<br />

from warranty coverage.<br />

5.1.1 Creating the Lab Environment<br />

To create satisfactory operating conditions in your lab environment, follow these<br />

guidelines:<br />

• Operate the ASXPRESS Rapid Sample Introduction System in a<br />

conventional lab environment where the temperature is 55°F to 85°F<br />

(13°C to 30°C), the humidity is 0% to 60% non-condensing and the<br />

unit is not exposed to excessive flammable or corrosive materials.<br />

• Avoid rough handling of the ASXPRESS Rapid Sample Introduction<br />

System. If possible, do not expose the system to vibration or shock.<br />

• Protect the system from long-term exposure to condensation,<br />

corrosive materials, solvent vapor, continual standing liquids, or any<br />

spills into the system cabinet. Exposures of this type can damage the<br />

internal mechanism components as well as the electronics.<br />

90