microstimulator with a 1:4 demultiplexer built on the electrode array results in 1024-<br />

pixel prosthesis.”, “For 1024 sites, 10 address bits are required. If the stimulus data for<br />

each driver is around 20 bits, this results in 50% increase in the data rate.”, “The chip<br />

consumes around 20 mW when delivering 600 A bi-phasic current pulses of 1 ms<br />

phase widths (anodic, cathodic and interphase delay) at a stimulation rate of 60 Hz.”<br />

[49]<br />

Electrical Stimulation in Isolated Rabbit Retina<br />

“Analysis of the data from this study predicts that a 50- m electrode may have enough<br />

safe charge capacity to evoke a retinal response. “, “With current electrode technology,<br />

more than one thousand 50- m-diameter electrodes could be placed in the macula (5-<br />

mm-diameter area around fovea) . Simulations of prosthetic vision, suggest that face<br />

recognition and reading will be possible with 1000 macular electrodes.”, “Overall,<br />

epiretinal electrodes and subretinal electrodes appear to have similar stimulus pulse<br />

requirements.” [229]<br />

A Biomimetic <strong>Retinal</strong> Stimulating Array<br />

“.. and in severe cases will lose most of their vision in the central 20º of the visual field.<br />

The macula is the part of the retina that anatomically corresponds to the central 20º of<br />

the visual field”… “When 32 × 32 electrodes were used, the recognition scores<br />

improved to over 80%.” “The electrodes should support 100 nC (100 μA for 1 ms)<br />

pulses without damaging either the electrodes or tissue.”…”Standard visual acuity<br />

testing demonstrated that 20/30 vision could be achieved using 625 pixels in the central<br />

1.7º of the visual field.”… “To accommodate a square arrangement, the electrode must<br />

be reduced to 93µm in order to achieve 1,000 macular electrodes.”<br />

[103]