<strong>Surgery</strong> Manual<br />
SCREW-CYLINDER-LINE Insertion<br />
Implant Bed Preparation for SCREW-CYLINDER-LINE Implants<br />
Preparation of the implant bed includes<br />
punch-marking the cortical bone, pilot<br />
drilling, pre-drilling, and form drilling.<br />
The pilot drilling and pre-drilling define<br />
the depth and axis of the implant bed<br />
and ensure tissue-conservative bone<br />
preparation by enlarging the diameter in<br />
small gradations.<br />
Pilot Drill and Pre-Drill Depth Stop<br />
The pilot drill and pre-drill have a maximum<br />
working length of 16 mm. The<br />
drilling depths of 9, 11, and 13 mm are<br />
laser-marked. An insertable depth stop<br />
limits the drilling depth to the selected<br />
depth of 9, 11, or 13 mm.<br />
Caution!<br />
The depth stops for pilot drills, and<br />
pre-drills five bladed cutter, shown<br />
here should be used only for<br />
implant bed preparation of ROOT-<br />
LINE, SCREW-CYLINDER-LINE, and<br />
CYLINDER-LINE Implants. These<br />
instruments are closely matched to<br />
one another.<br />
96<br />
Drill Extension<br />
A drill extension is available to prevent<br />
resting of the angled handpiece on the<br />
dental crown during preparation of the<br />
implant bed adjacent to elongated teeth.<br />
A longer irrigation tube is required for<br />
the internal irrigation system.<br />
Drill extension<br />
Depth stop for pilot drills and<br />
pre-drills