<strong>Surgery</strong> Manual<br />

SCREW-CYLINDER-LINE Insertion<br />

Pre-drilling Ø 3.3 mm<br />

Pre-drilling Ø 3.6 mm<br />

100<br />

Drilling depth check<br />

after pre-drilling Ø 3.3 mm<br />

Drilling depth check<br />

after pre-drilling Ø 3.6 mm<br />

Parallelism/axis check after<br />

pre-drilling Ø 3.3 mm<br />

Parallelism/axis check after<br />

pre-drilling Ø 3.6 mm<br />

Note<br />

If a simultaneous implantation into the<br />

augmented sinus is to be performed,<br />

pre-drilling continues only until penetration<br />

of the sinus cortical bone is felt.<br />

The use of a 9-mm depth stop is for<br />

demonstration purposes only since this<br />

configuration is also universally applicable.<br />

During final form drilling, the shortest<br />

possible form drill must be used to<br />

shape the Bioseal Bevel as far as the<br />

depth stop.