SCREW-CYLINDER-LINE Insertion<br />
Pre-drilling Ø 3.3 mm<br />
Pre-drilling Ø 3.6 mm<br />
100<br />
Drilling depth check<br />
after pre-drilling Ø 3.3 mm<br />
Drilling depth check<br />
after pre-drilling Ø 3.6 mm<br />
Parallelism/axis check after<br />
pre-drilling Ø 3.3 mm<br />
Parallelism/axis check after<br />
pre-drilling Ø 3.6 mm<br />
Note<br />
If a simultaneous implantation into the<br />
augmented sinus is to be performed,<br />
pre-drilling continues only until penetration<br />
of the sinus cortical bone is felt.<br />
The use of a 9-mm depth stop is for<br />
demonstration purposes only since this<br />
configuration is also universally applicable.<br />
During final form drilling, the shortest<br />
possible form drill must be used to<br />
shape the Bioseal Bevel as far as the<br />
depth stop.