1 Surgery CAMLOG Compendium
1 Surgery CAMLOG Compendium
1 Surgery CAMLOG Compendium
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Manual tapping is performed with the<br />
tap adapter for SCREW-LINE taps and<br />
the locked torque wrench. Make sure<br />
that the implant bed axis orientation is<br />
maintained during insertion and removal<br />
of the tap. The limit for insertion of the<br />
tap is the upper edge of the cutting<br />
blade. See also the instructions for the<br />
torque wrench on page 151.<br />
Caution!<br />
The SCREW-LINE taps must not<br />
be used for ROOT-LINE or SCREW-<br />
CYLINDER-LINE Implants.<br />
Locked torque wrench<br />
Tap Adapter, short / long<br />
<strong>Surgery</strong> Manual<br />
SCREW-LINE Insertion<br />
53